Technical FAQs

Ask a Question

What is the part number for the replacement for a M5IAI?

Issue:
Are repair parts available for the Medical panel indicator M5IAI?

Product Line:
Medical products sold and supported in the USA.

Environment:
Selection/maintenance

Cause:
Replacement

Resolution:
The M5IAI is now obsolete with no repair parts available. In order to replace the M5IAI with the current style product, the Line Isolation Monitor (LIM) will also need to be replaced. See below for the following options:
1. IG6M - LIM
2. 4800IG6CBKTVM - connector plate, harness, and mounting brackets
3. IG62000CBMST0614 - Only if the have the existing legacy box  530087BB OR
4. IG62000CBMG4 - mounts in an electrical 4 gang boxReplacement may require selection of additional components.
5. Please refer to the bulletin below for repair parts  
Should you have additional questions, please contact Technical Support @ 1.888.778.2733 for further assistance.


 
Was this helpful?
What can we do to improve the information ?