What is the recommended number of conductors and conduit size for 120V branch circuits in a hospital isolated power system?

Issue:
Installation/startup
Product Line:
Hospital Isolated Power Systems
Environment:
NFPA99, NEC517.160
Cause:
It is necessary to complete the installation of a hospital isolated power system in a manner which will comply with special requirements and pass required testing and certification.
Resolution:
3/4" conduit is recommended with no more than 6 conductors per conduit to help minimize leakage.